For full functionality of this publication it is necessary to enable Javascript.

Click here to see instructions how to enable JavaScript in your web browser.


<--

Analysis of radioactive iodine-129 using MS/MS with O2 reaction mode

Yasuyuki Shikamori, Kazumi Nakano and Naoki Sugiyama
Agilent Technologies, Japan

Keywords

radionuclide, iodine, 129I, environmental, nuclear, xenon, NIST 3231 Level I and II, abundance sensitivity, oxygen on-mass

Introduction

Iodine-129 is a long-lived radionuclide
(half-life of 15.7 My) which has been released into the environment as a result of human activities such as nuclear weapons testing, accidents at nuclear power plants and especially by emissions from spent nuclear fuel reprocessing plants. The determination of iodine-129 in environmental samples is very difficult by ICP-MS due to the element’s relatively low sensitivity, the very low concentrations at which 129I must be determined, relative to potentially high levels of 127I, the high background caused by 129Xe impurities in the argon plasma gas, and possible polyatomic interference from 127IH2+. Iodine analysis is further complicated by the fact that it is rapidly volatilized from samples prepared using the acid digestions that are normal for ICP-MS analysis, so an alternative, alkaline sample solubilization and stabilization strategy is required. The isobaric interference from 129Xe+ can be significantly reduced using ICP-QMS with an Octopole Reaction Cell operated in O2 reaction mode, resulting in a measured ratio for 129I/127I of 10-7  in NIST 3231 SRM Level I[1] However, the problem of potential overlap due to tailing from 127I and 127IH remains, as the relative abundance of the 129I to 127I will typically exceed 10-7, which is of the same order as the abundance sensitivity (ability to separate adjacent peaks) of quadrupole ICP-MS (ICP‑QMS). In order to overcome these challenges, ICP‑QQQ operating in MS/MS mode with O2 reaction gas was applied to determine ultratrace levels of iodine-129 in aqueous samples.

Experimental

Instrumentation: Agilent 8800 #100.

Plasma conditions: Preset plasma/Low matrix.

Ion lens tune: Soft extraction tune:
Extract 1 = 0 V, Extract 2 = -190 V.

CRC conditions: O2 gas at 0.8 mL/min,
Octopole bias = -18 V and KED = -1.5V. MS/MS O2 on-mass mode was applied to measure iodine-127 and iodine-129
(Q1 = Q2 = 127 for iodine-127,
Q1 = Q2 = 129 for iodine-129). 

Reference materials and calibration standards: Calibration standards were prepared by diluting 129I isotopic standards NIST SRM 3231 Level I and II
(NIST, Gaithersburg MD, USA) with
0.5% TMAH in deionized water. The
Level I Certified Value for
129I/127I = 0.981x10-6 ± 0.012x10-6,
Level II = 0.982x10-8 ± 0.012x10-8. These reference materials were used to check the calibration linearity of the iodine isotopes and to validate the isotopic ratio of iodine-129 and iodine-127. 

Results and discussion

Optimization of oxygen cell gas flow

The oxygen gas flow rate was optimized by varying the O2 flow over the full range of the mass flow controller (0–1.12 mL/min), while monitoring the 127,129I signal and blank intensity, as shown in Figure 1. As the flow rate of O2 increases, the background signal (due to 129Xe) at m/z = 129 decreases rapidly, and the iodine signal remains high, dramatically improving the DL for 129I.

 

Abundance sensitivity 

Scan spectra over the mass range 127 to 129, covering both 127I and 129I, were acquired for the two SRMs, NIST 3231 Level I and II, using the Agilent 8800 ICP-QQQ in MS/MS on-mass mode with O2 reaction gas. The overlaid spectra are shown in Figure 2. Excellent abundance sensitivity can be seen, with the sides of the intense (>109 cps) 127I peak reaching baseline with no tailing of 127I+ or 127IH+ on 129I+

 

Calibration curves for 127I and 129

In order to check the linearity of both iodine isotopes, different concentration solutions of NIST 3231 SRM Level I were prepared in 0.5% TMAH and analyzed as calibration standards, as shown in Figure 3. The BECs for 127I and 129I were 2.9 µg/L and 0.04 ng/L respectively, and the detection limits (3σ, n=10) were 0.26 µg/L for 127I and 0.07 ng/L for 129I.

 

Analysis of NIST 3231 SRM Level I and Level II 

The 129I/127I ratio in 10x diluted
NIST 3231 SRM Levels I
(129I/127I = 0.981 x 10-6) and
II (129I/127I = 0.982 x 10-8) was measured using ICP-QQQ in MS/MS on-mass mode with O2 cell gas. The results are summarized in Table 1. After subtracting the 129I blank, the measured 129I/127I ratio of NIST 3231 SRM Levels I and II corresponded well with the certified values of 0.981x10-6 and 0.982x10-8 respectively. The good agreement with the certified ratio indicates that the potential interference of 127IH2+ on 129I+ is completely removed by O2 reaction with MS/MS mode. 

Reference

  1. The ultratrace determination of iodine 129 in aqueous samples using the
    7700x ICP-MS with oxygen reaction mode, Agilent application note, 5990-8171EN. 

More information

  1. The ultratrace determination of iodine 129 using the Agilent 8800 Triple Quadrupole ICP MS in MS/MS mode, Agilent application note, 5991-0321EN.